Vergleich

ELISA Kit for Secretagogin (SCGN)

Hersteller Cloud-Clone
Kategorie
Typ Elisa-Kit
Specific against Human
Menge 96T
ArtNr SEG850Hu-96T
Targets CALB1, SCGN
eClass 6.1 32160605
eClass 9.0 32160605
Lieferbar
Quantity options
Detection range
31.2-2, 000pg/mL
Organism species
Homo sapiens (Human)
Sensitivity
The minimum detectable dose of this kit is typically less than 12.7pg/mL
Sample type
Serum, plasma, tissue homogenates and other biological fluids
Assay length
3h
Method
Double-antibody Sandwich
Specificity

This assay has high sensitivity and excellent specificity for detection of Secretagogin (SCGN).
No significant cross-reactivity or interference between Secretagogin (SCGN) and analogues was observed.

Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level Secretagogin (SCGN) were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level Secretagogin (SCGN) were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12%

Stability

The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.

Assay procedure summary
1. Prepare all reagents, samples and standards;
2. Add 100µ L standard or sample to each well. Incubate 2 hours at 37° C;
3. Aspirate and add 100µ L prepared Detection Reagent A. Incubate 1 hour at 37° C;
4. Aspirate and wash 3 times;
5. Add 100µ L prepared Detection Reagent B. Incubate 30 minutes at 37° C;
6. Aspirate and wash 5 times;
7. Add 90µ L Substrate Solution. Incubate 10-20 minutes at 37° C;
8. Add 50µ L Stop Solution. Read at 450nm immediately.
Test principle
The test principle applied in this kit is Sandwich enzyme immunoassay. The microtiter plate provided in this kit has been pre-coated with an antibody specific to Secretagogin (SCGN). Standards or samples are then added to the appropriate microtiter plate wells with a biotin-conjugated antibody specific to Secretagogin (SCGN). Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added, only those wells that contain Secretagogin (SCGN), biotin-conjugated antibody and enzyme-conjugated Avidin will exhibit a change in color. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Secretagogin (SCGN) in the samples is then determined by comparing the O.D. of the samples to the standard curve.
Research Area
Metabolic pathway;

Hinweis: Die dargestellten Informationen und Dokumente (Bedienungsanleitung, Produktdatenblatt, Sicherheitsdatenblatt und Analysezertifikat) entsprechen unserem letzten Update und sollten lediglich der Orientierung dienen. Wir übernehmen keine Garantie für die Aktualität. Für spezifische Anforderungen bitten wir Sie, uns eine Anfrage zu stellen.

Alle Produkte sind nur für Forschungszwecke bestimmt. Nicht für den menschlichen, tierärztlichen oder therapeutischen Gebrauch.

Menge: 96T
Lieferbar: In stock
lieferbar

Lieferung vsl. bis 30.05.2024 

Vergleichen

Auf den Wunschzettel

Angebot anfordern

Lieferzeit anfragen

Technische Frage stellen

Bulk-Anfrage stellen

Fragen zum Produkt?
 