Question 1: When performing the polymerization assay using pyrene-labeled actin (Cat. #AP05), the fluorescence signal of the globular actin in G-buffer increased over time, so that the 0 value was already rather high.
Answer 1: After preparing actin at a concentration of 0.4 mg/ml for polymerization (from either fresh stocks or frozen aliquots), it is important to incubate the actin on ice for 1 hour to depolymerize any actin polymers that might have formed during storage. For an even cleaner sample of G-actin, the samples can be centrifuged (100, 000 x g) for 60 min to absolutely insure the removal of residual actin oligomers and nucleating centers. After centrifugation, use the top 80% of the supernatant as your G-actin stock. This will eliminate any changes in fluorescence signal associated with G-actin.
Question 2: During my polymerization reaction, the growth phase and steady-state levels of F-actin produced are lower than what is shown in the manual. Why?
Answer 2: One of the most common causes of reduced fluorescence intensity signal is non-optimal machine settings. Please check that the machine&rsquo, s settings are similar to these:
Measurement type: Kinetic 120 cycles, 60 sec interval time
Fluorescence wavelengths: Ex. 350 or 360 +/- 20 nm
Em. 407 or 410 +/- 10 nm or 420 +/- 20 nm
Gain: 100 (on a scale of 0-120, where 120 is the highest)
Reads per well: 1
Fluorescence reading from: Top
Also, when samples are not being measured in the fluorimeter, the excitation beam should be shuttered to prevent the beam from bleaching the pyrene fluorescence. When measuring the fluorescence, the shutter should be opened for no longer than 7 sec every 30 sec for 1 h.
Hinweis: Die dargestellten Informationen und Dokumente (Bedienungsanleitung, Produktdatenblatt, Sicherheitsdatenblatt und Analysezertifikat) entsprechen unserem letzten Update und sollten lediglich der Orientierung dienen. Wir übernehmen keine Garantie für die Aktualität. Für spezifische Anforderungen bitten wir Sie, uns eine Anfrage zu stellen.
Alle Produkte sind nur für Forschungszwecke bestimmt. Nicht für den menschlichen, tierärztlichen oder therapeutischen Gebrauch.
AP05-B.pdf
AP05-B-datasheet.pdf
AP05-B-safety-datasheet.pdf
Vergleichen
Auf den Wunschzettel
Angebot anfordern
Lieferzeit anfragen
Technische Frage stellen
Bulk-Anfrage stellen
sales@hoelzel.de
« Zurück
Pünktliche Lieferung zu Ihrem Wunschtermin und bei jeder 5. Bestellung 15% sparen.
Passwort vergessen?
Noch nicht registriert? Konto hier erstellen!