Motilin (MTL) BioAssay™ ELISA Kit (Human) is a competitive inhibition immunoassay for in vitro quantitative measurement of MTL in human serum, plasma and other biological fluids.

Detection Range:
74.07-6000pg/ml

Sensitivity:
29.64pg/ml:


Intra-Assay CV:
<10%

Inter-Assay CV:
<12%

Kit Components:
*026884A: Microtiter Strips, 1x96 wells
*026884B: Standard, 2x1 vial
026884C: Standard Diluent, 1x20ml
*026884D: Detection Reagent A (green), 1x120ul
*026884E: Detection Reagent B (red), 1x120ul
026884F: Assay Diluent A, 1x12ml
026884G: Assay Diluent B, 1x12ml
026884H: TMB Substrate, 1x9ml
026884K: Stop Solution, 1x6ml
026884L: Wash Buffer, 30X, 1x20ml

Precaution:
The Stop Solution (026884K) included for use in this kit is an acid solution. Wear eye, hand, face, and clothing protection when using this material.

Storage and Stability:
Store *026884A, *026884B, *026884D and *026884E at -20°C. Store all the other components at 4°C. Unused kit is stable for 6 months after receipt. Once kit components are opened, it is highly recommended to use remaining reagents within 1 month provided this is within the expiration date of the kit. For maximum recovery of product, centrifuge the original vials after thawing and prior to removing the cap.

Assay Procedure Summary:
1. Prepare all reagents, samples and standards.
2. Add 50ul standard or sample to each well, then add 50ul prepared Detection Reagent A immediately. Shake and mix. Incubate 1 hour at 37°C.
3. Aspirate and wash 3 times.
4. Add 100ul prepared Detection Reagent B. Incubate 30 minutes at 37°C.
5. Aspirate and wash 5 times.
6. Add 90ul Substrate Solution. Incubate 15-25 minutes at 37°C.
7. Add 50ul Stop Solution. Read absorbance at 450nm immediately.
Shelf Life
6 months

Hinweis: Die dargestellten Informationen und Dokumente (Bedienungsanleitung, Produktdatenblatt, Sicherheitsdatenblatt und Analysezertifikat) entsprechen unserem letzten Update und sollten lediglich der Orientierung dienen. Wir übernehmen keine Garantie für die Aktualität. Für spezifische Anforderungen bitten wir Sie, uns eine Anfrage zu stellen.

Alle Produkte sind nur für Forschungszwecke bestimmt. Nicht für den menschlichen, tierärztlichen oder therapeutischen Gebrauch.

Menge: 96 tests
Lieferbar: In stock
lieferbar

Vergleichen

Auf den Wunschzettel

Angebot anfordern

Lieferzeit anfragen

Technische Frage stellen

Bulk-Anfrage stellen

Fragen zum Produkt?
 
Schließen