Comparison

RhoA G-LISA Activation Assay (colorimetric)

Manufacturer Cytoskeleton
Category
Type Assay
Specific against other
Amount KIT 96 assays
Item no. BK124
eClass 6.1 32161000
eClass 9.0 32161000
Available
Additional info

Product Uses Include

  • Rho signaling pathway studies
  • Rho activation assays with primary cells
  • Studies of Rho activators and inactivators
  • Rho activation assays with limited material
  • High throughput screens for Rho activation

Introduction
The G-LISA Rho activation assays are ELISA based Rho activation assays with which you can measure Rho activity in cells in less than 3 h. BK124 is very sensitive and has excellent accuracy between duplicate samples. For a more detailed introduction on G-LISA assays and a listing of other available G-LISA kits, see our main G-LISA page. The BK124 Rho activation assay kit measures the level of GTP-loaded RhoA only in cells. The level of activation is measured with absorbance set at 490nm. For a kit to measure RhoA activation with luminescence detection, see Cat. # BK121.

See G-LISA FAQs tab on our G-LISA page for more details.

Kit contents
The kit contains sufficient reagents to perform 96 RhoA activation assays. Since the Rho-GTP affinity wells are supplied as strips and the strips can be broken into smaller pieces, each kit can be used for anywhere from one to multiple assays. The following components are included in the kit:

  1. Rho-GTP affinity wells (12 strips of 8 wells)
  2. Lysis buffer
  3. Binding buffer
  4. Antigen presenting buffer
  5. Wash buffer
  6. Antibody dilution buffer
  7. Anti-RhoA antibody
  8. HRP-labeled secondary antibody
  9. Positive control RhoA protein
  10. Protease inhibitor cocktail (Cat. # PIC02)
  11. Absorbance detection reagents
  12. Precision Red Advanced protein assay reagent (Cat. # ADV02)
  13. Manual with detailed protocols and extensive troubleshooting guide

Equipment needed

  1. 96-well plate spectrophotometer capable of reading 490 nm wavelength
  2. Multichannel or multidispensing pipettor
  3. Orbital microplate shaker capable of at least 200 rpm shaking (400 rpm is optimal)

Example results
Serum starved Swiss 3T3 cells were stimulated with the Rho activating compound calpeptin and RhoA activation was measured with the G-LISA method (Figures 1 and 2)

Figure 1. RhoA activation by calpeptin measured by G-LISA kit BK124. Swiss 3T3 (mouse) cells were serum starved for 24 h and treated with calpeptin (Cal; 0.1 mg/ml for 30 min) or DMSO carrier only (SS). 10 ug of cell lysates were subjected to the G-LISA assay. Absorbance was read at 490 nm.

Figure 2. Rho activity measured in Swiss 3T3 cells treated with the Cell Permeable Rho Inhibitor (CT04) using the RhoA G-LISA Activation Assay (Cat.# BK124). Serum starved Swiss 3T3 fibroblasts were untreated (no CT04) or treated with 0.20, 0.50 and 2.0 ug/ml of CT04 for 4h in serum free medium at 37C, then activated with 100ug/ml calpeptin for 10min. Cells were then lysed and RhoA activity was measured by the RhoA G-LISA Activation Assay (Cat.# BK124). Note: At 2.0 ug/ml CT04 for 4h results in almost complete (90%) inhibition of RhoA activity.

Delivery Time
1-2 Weeks
Shipping Temp.
AT
Storage on Arrival
4C
faqs

Question 1:  Can I detect isoforms other than RhoA, Rac1 or RalA with these G-LISA activation assays?

Answer 1:  Yes, the RhoA G-LISA (Cat. # BK124), Rac1 G-LISA (Cat. # BK128) and RalA G-LISA (Cat. # BK129) can be used to detect RhoB or RhoC, Rac 2 or Rac3 or RalB, respectively.  The capture proteins that the wells have been coated with bind all of the isoforms of the respective GTPase.  The specificity of signal is conferred by the specificity of the monoclonal primary antibody utilized.  Use of an isoform-specific monoclonal antibody allows detection of other Rho family isoforms.  Please see this citation for an example of this modified procedure (Hall et al., 2008. Type I Collagen Receptor (&alpha, 2&beta, 1) Signaling Promotes Prostate Cancer Invasion through RhC GTPase. Neoplasia. 10, 797&ndash, 803). 

Basically the researcher would test their specific monoclonal antibody in a western blot first to prove specificity to the alternative isoform of interest.  For example, load RhoA and C for negative controls when testing a RhoB monoclonal antibody.  Then the researcher would use 1:50, 1:200 and 1:500 dilutions of their monoclonal antibody on duplicate cell extracts of activated and control state samples. The researcher would then choose the dilution of monoclonal antibody which gave them the highest ratio of activated:control state.

A simple activated/control state pair of extracts can be made by growing cells to 50% confluence in serum containing media, washing twice with PBS, preparing lysate and aliquoting and freezing   samples in liquid nitrogen.  With one aliquot, defrost and let stand at room temperature for 60 min to degrade the activated signal to a low basal signal, which will be the control state.   The untreated sample (2nd aliquot) will be considered &ldquo, activated&rdquo, which most serum grown cells are.

 

Question 2:  How many cell culture plates can I process at one time during the lysis step?

Answer 2:  We recommend that from the point at you add lysis buffer to the plate on ice to aliquoting and snap-freezing the lysate samples in liquid nitrogen, no more than 10 min are allowed to elapse.  After 10 min on ice, we find that GTP bound to GTPases (activated GTPases) undergoes rapid hydrolysis.  Rapid processing at 4°, C is essential for accurate and reproducible results.  The following guidelines are useful for rapid lysis of cells.

  Washing

a.  Retrieve culture dish from incubator, immediately aspirate out all of the media and place firmly on ice.

b.  Immediately rinse cells with an appropriate volume of ice cold PBS (for Cdc42 activation, skip this step and simply aspirate the media) to remove serum proteins.

c.  Aspirate off all residual PBS buffer. This is essential so that the Lysis Buffer is not diluted. Correct aspiration requires that the culture dish is placed at a steep angle on ice for 1 min to allow excess PBS to collect in the vessel for complete removal.  As noted, the time period between cell lysis and addition of lysates to the wells is critically important. Take the following precautions:

        1.  Work quickly.

        2.  Keeping solutions and lysates embedded in ice so that the temperature is below 4°, C. This helps to minimize changes in signal over time.

        3.  We strongly recommend that cell lysates be immediately frozen after harvest and clarification. A sample of at least 20 &mu, l should be kept on ice for protein concentration measurement. The lysates must be snap frozen in liquid nitrogen and stored at -70°, C. Lysates should be stored at -70°, C for no longer than 30 days.

        4.  Thawing of cell lysates prior to use in the G-LISA assay should be in a room temperature water bath, followed by rapid transfer to ice and immediate use in the assay.

 

If you have any questions concerning this product, please contact our Technical Service department at infohoelzel.com.

Note: The presented information and documents (Manual, Product Datasheet, Safety Datasheet and Certificate of Analysis) correspond to our latest update and should serve for orientational purpose only. We do not guarantee the topicality. We would kindly ask you to make a request for specific requirements, if necessary.

All products are intended for research use only (RUO). Not for human, veterinary or therapeutic use.