Comparison

TIGIT ELISA Kit (Human)

Item no. OKCD01082
Manufacturer AVIVA Systems Biology
Amount 96 Wells
Category
Type Elisa-Kit
Applications ELISA
Specific against Human (Homo sapiens)
Sensitivity < 0.133 ng/mL
ECLASS 10.1 32160605
ECLASS 11.0 32160605
UNSPSC 41116126
Alias T-cell immunoreceptor with Ig and ITIM domains;V-set and immunoglobulin domain containing 9;V-set and immunoglobulin domain-containing protein 9;V-set and transmembrane domain containing 3;V-set and transmembrane domain-containing protein 3;VSIG9;VSIG9,VSTM3;VSTM3;Washington University cell adhesion molecule;WUCAM.
Shipping Condition Cool pack
Available
Manufacturer - Type
ELISA Kit
Manufacturer - Applications
Enzyme-linked Immunosorbent assay-Sandwich
Manufacturer - Category
Root Catalog/Products/ELISA Kits, Root Catalog/Products/Kit
Shipping Temperature
Wet Ice
Gene symbol
TIGIT
Gene Fullname
T cell immunoreceptor with Ig and ITIM domains
Protein size
244
Reconstitution and storage
2°C to 8°C|-20°C
Description of target
Binds with high affinity to the poliovirus receptor (PVR) which causes increased secretion of IL10 and decreased secretion of IL12B and suppresses T-cell activation by promoting the generation of mature immunoregulatory dendritic cells.
Nucleotide accession_num
NM_173799.3
Protein accession_num
NP_776160
Protein name
T-cell immunoreceptor with Ig and ITIM domains
Manufacturer - Specificity
This assay has high sensitivity and excellent specificity for detection of T-Cell Immunoreceptor With Ig And ITIM Domains Protein (TIGIT).No significant cross-reactivity or interference between T-Cell Immunoreceptor With Ig And ITIM Domains Protein (TIGIT) and analogues was observed.
Assay info
Assay Methodology: Quantitative Sandwich Immunoassay
Kit Components
Component: Amount
Anti-TIGIT Microplate: 96 Wells (12 x 8 Well Strips)
TIGIT Lyophilized Standard: 2 x 20 ng
100X Biotinylated TIGIT Detector Antibody: 120 uL
Avidin/HRP Conjugate: 120 uL
Standard Diluent: 1 x 20 mL
Detector Antibody Diluent: 1 x 12 mL
Conjugate Diluent: 1 x 12 mL
30X Wash Buffer: 1 x 20 mL
TMB Substrate: 1 x 9 mL
Stop Solution: 1 x 6 mL
Kit detection
Colorimetric
Kit duration
3h
Kit linearity
Sample - 1:2 Dilution - 1:4 Dilution - 1:8 Dilution - 1:16 Dilution
serum(n=5) - 78-89% - 92-101% - 92-105% - 89-103%
EDTA plasma(n=5) - 88-105% - 79-104% - 87-96% - 80-101%
heparin plasma(n=5) - 78-95% - 96-103% - 87-101% - 78-94%
Kit principle
The test principle applied in this kit is Sandwich enzyme immunoassay. The microtiter plate provided in this kit has been pre-coated with an antibody specific to T-Cell Immunoreceptor With Ig And ITIM Domains Protein (TIGIT). Standards or samples are then added to the appropriate microtiter plate wells with a biotin-conjugated antibody specific to T-Cell Immunoreceptor With Ig And ITIM Domains Protein (TIGIT). Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added, only those wells that contain T-Cell Immunoreceptor With Ig And ITIM Domains Protein (TIGIT), biotin-conjugated antibody and enzyme-conjugated Avidin will exhibit a change in color. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm +/- 10nm. The concentration of T-Cell Immunoreceptor With Ig And ITIM Domains Protein (TIGIT) in the samples is then determined by comparing the O.D. of the samples to the standard curve.
Kit recovery
Matrix - Recovery Range (%) - Average (%)
serum(n=5) - 78-91 - 86
EDTA plasma(n=5) - 93-104 - 99
heparin plasma(n=5) - 92-101 - 97
Kit reproducibility
Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level T-Cell Immunoreceptor With Ig And ITIM Domains Protein (TIGIT) were tested 20 times on one plate, respectively. Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level T-Cell Immunoreceptor With Ig And ITIM Domains Protein (TIGIT) were tested on 3 different plates, 8 replicates in each plate. CV(%) = SD/meanX100 Intra-Assay: CV<10% Inter-Assay: CV<12%
Sample Type
Serum, plasma, tissue homogenates and other biological fluids

Note: The presented information and documents (Manual, Product Datasheet, Safety Datasheet and Certificate of Analysis) correspond to our latest update and should serve for orientational purpose only. We do not guarantee the topicality. We would kindly ask you to make a request for specific requirements, if necessary.

All products are intended for research use only (RUO). Not for human, veterinary or therapeutic use.

Amount: 96 Wells
Available: In stock
available

Compare

Add to wishlist

Get an offer

Request delivery time

Ask a technical question

Submit a bulk request

Questions about this Product?
 
Close