Comparison

Estradiol (E2) BioAssay™ ELISA Kit

Item no. USB-024887-96Tests
Manufacturer United States Biological
Amount 96 tests
Quantity options 96 tests 96 tests
Category
Type Elisa-Kit
Specific against other
ECLASS 10.1 32160605
ECLASS 11.0 32160605
UNSPSC 41116126
Shipping Condition Cool pack
Available
Manufacturer - Category
Kits and Assays / Kits-ELISA (Hormones & Steroids), BioAssay™
Shipping Temperature
Blue Ice
Storage Conditions
4°C/-20°C
EU Commodity Code
38220000
Description
The Estradiol (E2) ELISA Kit is a competitive inhibition immunoassay for the in vitro quantitative measurement of E2 in serum, plasma, tissue homogenates, cell lysates, cell culture supernatants and other biological fluids.

Detection Range:
12.35-1, 000pg/ml

Sensitivity:
<4.75pg/ml

Precision:
Intra-Assay CV: <10%
Inter-Assay CV: <12%

Test Principle:
This assay employs the competitive inhibition enzyme immunoassay technique. A monoclonal antibody specific for E2 has been pre-coated onto a microplate. A competitive inhibition reaction occurs between biotin-labeled E2 and unlabeled E2 (standards or samples) with the pre-coated antibody specific for E2. After incubation the unbound conjugate is washed off. Next, avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. The amount of bound HRP conjugate is inversely proportional to the concentration of E2 in the sample. After addition of the substrate solution, the intensity of color developed is inversely proportional to the concentration of E2 in the sample.

Kit Components:
*024887A: Microtiter Plate, 96 wells, Pre-coated, ready to use
*024887B: Standard, 2x1vial
024887C: Standard Diluent, 1x20ml
*024887D: Detection Reagent A, 1x120ul
*024887E: Detection Reagent B, 1x120ul
024887F: Assay Diluent A, 1x12ml
024887G: Assay Diluent B, 1x12ml
024887H: TMB Substrate, 1x9ml
024887K: Stop Solution, 1x6ml
024887L: Wash Buffer, 30X, 1x20ml

Storage and Stability:
Store *024887A, *024887B, *024887D and *024887E at -20°C. Store all the other components at 4°C. Unused kit is stable for 6 months. Once kit components are opened, it is highly recommended to use remaining reagents within 1 month provided this is within the expiration date of the kit. For maximum recovery of product, centrifuge the original vials after thawing and prior to removing the cap.

Assay Procedure Summary:
1. Prepare all reagents, samples and standards.
2. Add 50ul standard or sample to each well, then add 50ul prepared Detection Reagent A immediately. Shake and mix. Incubate 1 hour at 37°C.
3. Aspirate and wash 3 times.
4. Add 100ul prepared Detection Reagent B. Incubate 30 minutes at 37°C.
5. Aspirate and wash 5 times.
6. Add 90ul Substrate Solution. Incubate 15-25 minutes at 37°C.
7. Add 50ul Stop Solution. Read absorbance at 450nm immediately.
Shelf Life
6 months

Note: The presented information and documents (Manual, Product Datasheet, Safety Datasheet and Certificate of Analysis) correspond to our latest update and should serve for orientational purpose only. We do not guarantee the topicality. We would kindly ask you to make a request for specific requirements, if necessary.