Comparison

IL-10 High Sensitivity Kit (Human)

Item no. OKDB00042
Manufacturer AVIVA Systems Biology
Amount 1 x 96 Wells
Category
Type Elisa-Kit
Specific against Human (Homo sapiens)
Sensitivity < 0.98 pg/mL
ECLASS 10.1 32160605
ECLASS 11.0 32160605
UNSPSC 41116126
Alias CSIF,TGIF,GVHDS,IL-10,IL10A
Shipping Condition Cool pack
Available
Manufacturer - Type
ELISA Kit
Manufacturer - Category
Root Catalog/Products/Kit
Shipping Temperature
Wet Ice
Molecular Weight
21 kDa
Gene symbol
IL-10
Gene Fullname
interleukin 10
Protein size
178
Reconstitution and storage
2 to 8C, Do not use past expiration date.
Description of target
The protein encoded by this gene is a cytokine produced primarily by monocytes and to a lesser extent by lymphocytes. This cytokine has pleiotropic effects in immunoregulation and inflammation. It down-regulates the expression of Th1 cytokines, MHC class II Ags, and costimulatory molecules on macrophages. It also enhances B cell survival, proliferation, and antibody production. This cytokine can block NF-kappa B activity, and is involved in the regulation of the JAK-STAT signaling pathway. Knockout studies in mice suggested the function of this cytokine as an essential immunoregulator in the intestinal tract. Mutations in this gene are associated with an increased susceptibility to HIV-1 infection and rheumatoid arthritis.
Nucleotide accession_num
NM_00572.2
Protein accession_num
NP_000563.1
Protein name
interleukin-10
Manufacturer - Specificity
Ten specificities were tested with concentrations higher than IL-10 curve concentrations. No cross reaction was observed for concentrations for IL-1alpha and beta, IL-2, IL-5, IL-6, IL-8, IL-12, TNF-alpha, IFN-gamma and mIL-10.
Assay info
Assay Methodology: Quantitative Sandwich ELISA
Kit duration
< 4 Hours
Kit linearity
Two spiked human serum with different levels of IL-10 were analysed at three serial two fold dilutions (1:2-1:16) with two replicates each. Recoveries ranged from 90% to 117% with an overall mean recovery of 101%.
Kit principle
Aviva Systems Biology IL-10 High Sensitivity ELISA Kit (Human) (OKDB00042) is designed for the highly sensitive in-vitro quantitative measurement of recombinant or natural IL-10 in human serum, plasma or cell culture media. The kit is based on standard sandwich enzyme-linked immuno-sorbent assay methods along with an additional detection process employing a polymerization step to amplify the number of detector bound HRP molecules. A monoclonal antibody specific for IL-10 has been pre-coated into wells of a 96-well plate. Standards or test samples are added to the wells and incubated. After washing, a biotinylated detector antibody specific for IL-10 is added, incubated and followed by washing. Streptavidin-HRP conjugate is then added, incubated and unbound conjugate is washed away. Amplification is achieved by adding a Biotin-Tyramine reagent, incubating and washing, followed by a second incubation with Streptavidin-HRP conjugate. An enzymatic reaction is then produced through the addition of TMB substrate which is catalyzed by HRP to generating a blue color product that changes yellow after adding acidic stop solution. The density of yellow coloration read by absorbance at 450 nm and is quantitatively proportional to the amount of sample captured in well.
Kit recovery
The spike recovery was evaluated by spiking two concentrations of recombinant IL-10 in human serum and culture media in three experiments. Recoveries ranged from 70% to 101% with an overall mean recovery of 85%.
Kit reproducibility
Four sera and one cell culture medium samples with various concentrations of IL-10 were tested for repeatability and reproducibility. Each assay was carried out with 3 replicates of each sample. Five independent assays were performed. The intra-assay and inter-assay coefficient of variation has been calculated to be 4.3% and 6.3% respectively.
Sample Type
Serum, Plasma, Tissue Homogenates , Cell Culture Supernates, Other Biological Fluids

Note: The presented information and documents (Manual, Product Datasheet, Safety Datasheet and Certificate of Analysis) correspond to our latest update and should serve for orientational purpose only. We do not guarantee the topicality. We would kindly ask you to make a request for specific requirements, if necessary.