Comparison

Influenza A H5N1 (Avian Flu) Hemagglutinin / HA ELISA Pair Set European Partner

Item no. SEK002-5
Manufacturer Sino Biological
Amount 5 Plates
Quantity options 15 Plates 5 Plates
Category
Type ELISA Pair Set
Applications ELISA
Specific against Flu viruses (Influenza)
Conjugate/Tag Biotin
Sensitivity The minimum detectable dose of H5N1 hemagglutinin (HA) was determined to be approximately 78.125 pg/ml. This is defined as at least three times standard deviations above the mean optical density of 10 replicates of the zero standard.
ECLASS 10.1 32160605
ECLASS 11.0 32160605
UNSPSC 41116126
Alias SEK002
Similar products HA
Available
Storage Conditions
Capture Antibody: Aliquot and store at -20C to -80C for up to 6 months from date of receipt. Avoid repeated freeze-thaw cycles.

Detection Antibody: Aliquot and store at -20C to -80C for up to 6 months from date of receipt. Avoid repeated freeze-thaw cycles.

Standard: Store lyophilized standard at -20C to -80C for up to 6 months from date of receipt. Aliquot and store the reconstituted Standard at -80C for up to 1 month. Avoid repeated freeze-thaw cycles.

Streptavidin-HRP: Store at 4C and protect it from prolonged exposure to light. DO NOT FREEZE! It is stable for up to 6 months from date of receipt.
Antigen molecule name
HA
Specificity
This H5N1 HA / Hemagglutinin ELISA Pair Set (SEK002) can quantificationally detect Anhui-HA (A/Anhui/1/2005), Indonesia-HA (A/Indonesia/5/2005), and Vietnam-HA (A/Vietnam/1194/2004). And it can also react with Turkey-HA (A/turkey/Turkey/1/2005) and Qinghai-HA (A/bar-headed goose/Qinghai/14/2008).
Capture Ab
0.5 mg/mL of mouse anti-H5N1 hemagglutinin monoclonal antibody. Dilute to a working concentration of 0.5 ug/mL in PBS before coating.
Detection Ab
0.5 mg/mL biotinylated rabbit anti-H5N1 hemagglutinin polyclonal antibody. Dilute to a working concentration of 1.5 ug/mL in detection antibody dilution buffer before use.
Standard
Each vial contains 150 ng of recombinant H5N1 hemagglutinin. Reconstitute with 1 mL detection antibody dilution buffer. A seven-point standard curve usi ng 2-fold serial dilutions in sample dilution buffer, and a high standard of 5000 pg/mL is recommended.
Principle of the test
The H5N1 (avian flu) hemagglutinin ELISA Pair Set is for the quantitative determination of hemagglutinin.
This ELISA Pair Set contains the basic components required for the development of sandwich ELISAs.
The Sino ELISA Pair Set is a solid phase sandwich ELISA (Enzyme-Linked Immunosorbent Assay). It utilizes a monoclonal antibody specific for hemagglutinin coated on a 96-well plate. Standards and samples are added to the wells, and any hemagglutinin present binds to the immobilized antibody. The wells are washed and a biotinylated rabbit anti-hemagglutinin polyclonal antibody is then added, producing an antibody-antigen-antibody "sandwich". To produces color in proportion to the amount of hemagglutinin present in the sample strepavidin-HRP and TMB substrate solution are loaded. The absorbances of the microwell are read at 450 nm.
Streptavidin-HRP
50 uL of streptavidin conjugated to horseradish-peroxidase. 1:2000 Dilution in detection antibody dilution buffer before use.

Note: The presented information and documents (Manual, Product Datasheet, Safety Datasheet and Certificate of Analysis) correspond to our latest update and should serve for orientational purpose only. We do not guarantee the topicality. We would kindly ask you to make a request for specific requirements, if necessary.

All products are intended for research use only (RUO). Not for human, veterinary or therapeutic use.

Amount: 5 Plates
Available: In stock
available

Compare

Add to wishlist

Get an offer

Request delivery time

Ask a technical question

Submit a bulk request

Questions about this Product?
 
Close